40
RBCBioscience
5. Press the V-axis downward until the tool can be put into the screw holes above the
pistons.
6. Screw off the four screws (2.5mm Hex wrench).
7. Disassemble the fault pistons and install new pistons (2.5mm Hex wrench).
8. Screw up the four screws (2.5mm Hex wrench).
9. Screw up the screws above the piston (2.5mm Hex wrench).
10. Reinstall the acrylic nameplate of the piston module and the LED lamp on the
machine.
11. After the above steps are finished, please perform “Piston Test“ to check all
processes are successful.