25
━ Reject Tip
Please select the Piercing Needles button.
1. Please prepare a soft cloth with 70% alcohol.
2. Please clean the Piercing needles by the cloth.
If the system is stopped from power failure or other
abnormal conditions and the tips are still stayed on
the pistons, you have to press Reject Tip to remove
the tips from the pistons.
Please select the Reject Tip button.
━ Clean Piercing Needles
2
1
1
Stop
Piercing Needles
Next Step
Prev
Maintenance
Update Program
Vender
Piercing Needles
System Status
Optic Test
Alarm History
Back
Maintenance
Reject Tip
Piston Test
UV
Heater Test
Replace O-Ring
Motor Test
Next
Back
Warning!
The tips of the Piercing Needles
are extremely sharp. Please
clean them carefully.
Summary of Contents for MagCore HF16 Plus
Page 1: ...MagCore Super HF16Plus EngineerInTraining manual ...
Page 34: ...33 Disassemble thebackservicelid Disassemble back service lid ...
Page 35: ...34 RBCBioscience Disassemblethe topservicelid Po ...
Page 63: ...62 RBCBioscience Maintenance Electricalcontrolandengineeringsoftware ...
Page 66: ...65 Mainboardconfiguration ...
Page 75: ...74 RBCBioscience PositionTeachingTools 200µlSPTip toolA toolB toolC PipetteTip ...
Page 87: ...86 RBCBioscience ...
Page 93: ...92 RBCBioscience ...
Page 94: ......
Page 95: ......