Instructions for Use MPR1 DATALOGGER
Manufacturer:
RAUMEDIC AG, Hermann-Staudinger-Strasse 2, 95233 Helmbrechts, Germany
Page 77 of 98
6.7 Pressure sensor has been removed
The pressure sensor PNo. was removed. Confirm this with OK. If you haven't removed the sensor,
the cause is most probably a sensor defect. Please check this and swap the sensor if needs be!
PNo. stands for pressure channel P1, P2 or P3, depending on which pressure channel is concerned.
6.8 Temperature sensor has been removed
The temperature sensor TNo. has been removed. Confirm this with OK. If you haven't removed the
sensor, the cause is most probably a sensor defect. Please check this and swap the sensor if needs
be!
TNo. stands for temperature channel T1 or T2, depending on which temperature channel is concerned.
6.9 ECG cable disconnected
The ECG cable has been removed. Confirm this with OK. If you have not removed the cable, it is most
likely a cable defect. Please check this and replace the cable if necessary.
6.10 Grave error A
Error message:
Grave error
A defect has been detected in the device settings, these will be reset to factory settings.
Once you have switched the device off and on again, you can continue working. Check
the settings (e.g. date, time, alarm limits).
Please follow the instructions ! It is imperative that the settings of the apparatus are checked !
Cause: RAM error
6.11 Grave error B
Error message:
Grave error
A grave error system defect (error number) has arisen. The device will no longer work.
Please switch the device off.
Note:
It is imperative to note down the error number before switching the apparatus off. The
most probable cause is a fault in the power supply of the apparatus. If the error is in direct
conjunction with an external event, such as the apparatus has dropped, defibrillation of
the patient, et cetera, the apparatus has to be switched on and off again. If the error has
disappeared, continue to operate the unit for the time being. If the error occurs again,
switch the unit off and get in touch with the after-sales service department.
6.12 Internal error – analogue output
Internal defect in chosen analogue output. Use other analogue output if needed.
Here the user cannot use the selected analogue output. A continuation of the logger operation without the
use of the analogue output concerned is possible at any time. The fault can be remedied by the after-sales
service department only.