The following table indicates how the different alarms are acoustically and optically noticed by the operator.
Alarm
condition
LED [6]
*1
LED [7]
*1
Display
Sound
none
green
off
normal
none
*2
sensor
off
yellow
„Sensor ?“ approx. 30 s. break
medium
priority
*3
off
yellow
value
flashes
approx. 20 s. break
high priority
red
red
value
flashes
approx. 7 s. break
Table 7: Alarm conditions
*1: See Table 1 in chapter 2.3 Contraindications
*2: Individual actions are indicated by individual short sounds.
*3: Physiological and technical alarm
4.9.7 Verifying the alarm system
The alarm system can be verified at any time. Please observe when verifying the acoustic signal that the
muting / reset key has not switched off the acoustic alarm permanently. To verify the alarm system, proceed
as described in chapter
Device diagnostics. Please observe when verifying the alarm system that
alarms of high and medium priority are tested according to the table rendered above.