Instructions for Use MPR1 DATALOGGER
Manufacturer:
RAUMEDIC AG, Hermann-Staudinger-Strasse 2, 95233 Helmbrechts, Germany
Page 9 of 98
2.2.6 Application in connection with electrosurgical units
Devices that meet the requirements of IEC 60601-1-2:
2007
(for differentiation, see Section
10 Electromagnetic compatibility) are contraindicated for simultaneous use with electrosurgical devices.
(see section 2.3 Contraindications).
Only devices
that fulfill the requirements of IEC 60601-1-2:
2014
are
suitable for simultaneous use with electrosurgical units. The following notes must be observed:
•
If the MPR1 is used together with an electric surgical instrument, then a measurement inaccuracy
according to chapt. 10.1.1 Accuracy under the influence of electromagnetic interference phenomena
can occur.
•
Before using electrosurgical units, check that the test transducers used are equipped with appropriate
protective measures against patient burns. If the sensors do not have suitable protective measures,
they must be disconnected from the device before electrosurgical units are used.
2.3 Contraindications
The MPR1 is contra-indicated for use in CT or MRI and must not be operated in a corresponding
environment.
Devices that meet the requirements of IEC 60601-1-2:2007 (for differentiation, see Section
10 Electromagnetic compatibility) are contraindicated for simultaneous use with electrosurgical devices.
The unit is not intended for use outside of the hospitals, such as in helicopters or in ambulances. The MPR1
is not intended for use in domestic settings.
2.4 Main operating functions
The main operating functions of the MPR1 are:
•
the installation of the unit, connect system components
•
the installation of the device, connect system components;
•
switch on the device;
•
connect the catheter;
•
selecting the recording mode;
•
connecting to a third-party system;
•
selecting the catheter zeroing;
•
determining the measurement location;
•
erasing the memory;
•
create patient file;
•
performing a simple functional pre-test of the alarm signals;
•
configure alarms;
•
deactivating the alarms;
•
measure;
•
display trend;
•
set storage mode;
•
monitor the temperature output value;
•
perform a simple functional pre-test of the medical thermometer;
•
switch off the device;
•
cleaning and disinfecting
The configuration of the unit is no main operating function.
2.5 Operating elements, connections, displays
The following two pages will render an overview of the most important operating elements and displays.
The unit has been fitted with two types of keys:
1 Keys with dedicated functions and
2 Keys with context-sensitive functions
This
document
is
for
information