![background image](http://html1.mh-extra.com/html/rands/fsh-series/fsh-series_quick-start-manual_3147226093.webp)
91
E-02
When measuring high powers, pay strict attention
to the following instructions to avoid personal
injury and to prevent the power sensor from being
destroyed:
Danger of skin burns and damage to the
instrument
Never exceed the permissible continuous power.
•
See diagram on the rear of the sensor for the
•
permissible continuous power.
Turn off the RF power to connect the sensor.
•
Screw the RF connectors tightly.
•
Operation
Connect the cable on the power sensor to
the power sensor socket on the R&S FSH and
screw it on. Insert the directional power sensor
between the source and the load.
Press the MODE key.
Press the POWER METER softkey.
The R&S FSH changes to power measurement
mode and opens the screen and the menu for
the directional power measurement. If no power
sensor is connected, no measured value is
displayed. If a power sensor is connected, the
R&S FSH establishes a connection to the power
sensor via the interface and, after a few seconds,
displays the connected power sensor type
(R&S FSH-Z14 or R&S FSH-Z44) as well as the
measured forward power and return loss of the
load.
GalaxyE_v1-10_3.indd 91
GalaxyE_v1-10_3.indd 91
08.04.2009 15:21:50
08.04.2009 15:21:50