31
E-02
The carrying handle at the top of the R&S FSH can
also be used to hang it from cabinet doors, for
example. The shape of the grip ensures that the
instrument does not fall off.
2.2.3
Switching on the Spectrum
Analyzer
The rechargeable lithium ion battery included
in the delivery must be inserted into the battery
compartment before the R&S FSH is switched on.
The battery compartment is situated at the lower
right corner of the instrument.
To open the battery
compartment, push the cover
downward by pressing a
fi nger against the recessed
grip (1). Then open the cover
upwards (2) and put the
rechargeable battery into the
R&S FSH as far as it will go
with the beveled side facing
upward and the grip on the
right side (3). To close the
cover, fold it downward (4)
and push it upward until it
clicks fully into place (5).
The R&S FSH can be operated with the AC
power supply included in the delivery or with the
replaceable battery. The lithium ion battery which
is supplied has a capacity of approximately 4 Ah
and permits up to three hours’ operation when
fully charged. The R&S FSH battery may not be
GalaxyE_v1-10_3.indd 31
GalaxyE_v1-10_3.indd 31
08.04.2009 15:21:40
08.04.2009 15:21:40