Alternatively, the start and stop frequency can
be input using the START FREQ and STOP FREQ
softkeys in the frequency menu.
3.7.1
Calibrate the R&S FSH for the
Vector Return Loss Measurement
Press the MEAS key.
Press the CALIBRATE softkey.
Select REFLECTION PORT 1 using the rotary
knob or the cursor keys.
Confi rm with ENTER or by pressing the
CALIBRATE softkey.
The R&S FSH displays a message requesting
termination of the measurement input into an open
circuit (Open).
Terminate the measurement input of the VSWR
bridge or the end of the measurement cable
into an open circuit.
Start open-circuit calibration by pressing
CONTINUE.
When the open-circuit calibration is done, the
R&S FSH outputs a message requesting the
termination of the measurement port into a short
circuit (SHORT).