The R&S FSH changes to power measurement
mode and opens the screen for the power
measurement. If a power sensor has not been
connected, no measured value is displayed. If a
power sensor has been connected, the R&S FSH
sets up a connection via the interface and, after a
few seconds, displays the measured power.
If there are any communication problems with
the power sensor, the R&S FSH outputs error
messages (sensor error: error number) indicating
the possible causes (see main manual).
To compensate for internal offset of the power
meter, it needs to be compensated before starting
the measurement.
Press the ZERO softkey.
The R&S FSH outputs a message telling you not to
apply any signals to the power meter while zeroing
is being performed.
Disconnect the power sensor from any signal
sources.
Start zeroing with the softkey CONTINUE.