PD 1309.6269.62-02
Connect the generator output with the 50
Ω
termination.
Press the CONTINUE softkey.
When calibration is over, TRANSM. VECTOR (CAL)
is displayed in the upper area of the screen to
indicate that the R&S FSH is vector-calibrated for
transmission measurement.
3.6.2 Calibrating the R&S FSH for Scalar
Measurement of the Transmission
Function
The following example shows a scalar
measurement of the transmission function. If the
option R&S FSH-K42 is installed, the measurement
must fi rst be switched to scalar (only for R&S FSH
models 1309.6000.24 or 1309.6000.28).
Select VECTOR to increase the accuracy
(for R&S FSH models 1309.6000.24 and
1309.6000.28 with option K42 (vector
network measurements) only). For
R&S FSH models 1309.6000.14 and
1309.6000.18 and without option K42 only
SCALAR can be selected.
GalaxyE_v1-10_3.indd 96
GalaxyE_v1-10_3.indd 96
08.04.2009 15:21:50
08.04.2009 15:21:50