After the calibration is fi nished, the R&S FSH
displays (CAL) at the top. This indicates that
the R&S FSH is calibrated for scalar refl ection
measurements.
Scalar refl ection measurements
Connect the device under test to the calibrated
measurement port.
The R&S FSH displays the return loss of the device
under test in dB:
The R&S FSH additionally offers to display the
refl ection in VSWR or as a refl ection coeffi cient.
Proceed as follows:
Select the FORMAT softkey.
Select VSWR or REFLECTION COEFFICIENT with
the rotary knob or the cursor keys (
∧
or
∨
).
Confi rm the selection with the ENTER key or the
FORMAT softkey.