Terminate the measurement input of the bridge
or the end of the measurement cable into a
short circuit.
Start the short-circuit calibration by pressing
CONTINUE.
In the third step of calibration, terminate the
measurement port into a 50
Ω
termination.
Terminate the measurement input of the bridge
or the end of the measurement cable into a
50
Ω
termination (LOAD)
Start termination calibration by pressing
CONTINUE.
When calibration is done, the R&S FSH outputs
the message “Calibration Done” for three seconds.
REFL. VECTOR (CAL) is displayed in the upper area
of the diagram. The R&S FSH is vector-calibrated
for refl ection measurements.
Vector refl ection measurements
Select a measurement within the FORMAT softkey
menu. A detailed description of the measurements
is given in the Oerating Manual in chapter
“Measuring the Refl ection Magnitude”
.