EB200 Manual
Operation
4052.2000.02 3.59
E-7
Inhibiting the operating elements at remote control
The message in the softkey area of the display is switched on and off by the remote control command
SYSTem:KLOCk ON|OFF
. At the same time the keyboard and the spinwheel are disabled. Now only
pure remote-control operation is possible.
In this mode, a DATA lamp comes on in the text window.
The test in the message window can be specified by REMOTE with a special command.
The Rohde & Schwarz default text is: KEYBOARD CONTROLLED BY REMOTE.
Inhibiting the display and operating elements at remote control
The display can be switched on and off by the remote-control command
DISPlay:ENABle ON|OFF
.
In this case, the entire front panel editing is disabled, providing additional processor capacity for remote
control and hardware control.
RECEIVER MODE:
Indicates the current mode of the unit set via remote-control
command
FREQuency:MODE
.
CW
Receiver monitors a frequency.
SWEEP
Receiver is in frequency sweep mode.
MSCAN
Receiver is in memory scan.
DSCAN
Receiver is in Digiscan mode.
FASTLEVCW
Receiver is in mode for fast triggered level measurement.
LIST
Receiver is in mode for fast level measurement in a frequency list.
The DSCAN mode can only be switched on if the SW option EB200DS (Digiscan) is fitted.
The FASTLEVCW and LIST modes can only be switched on if the SW option EB200CM (Coverage
Measurement) is fitted.
The text in the 2
nd
line of the message window can be pre-defined by using the remote-control
command
DISPlay:ENABle:LABel „String with max 24 chars“
. The text pre-defined by
Rohde & Schwarz is "CONTROLLED BY REMOTE".