EB200 Manual
SENSe subsystem VOLTage
4052.2000.02 4.94
E-7
. :VOLTage:AC:RANGe
. . [:UPPer]<numeric_value>UP|DOWN|MINimum|MAXimum
This command is used in the DSCAN mode to set the reference level. In order to avoid overdriving of
the receiver, the reference level is set to the highest expected positive signal level.
Parameters:
<numeric_value>
reference level in dBµV
UP|DOWN
|increase | decrease of the reference level by the value that was
set by command
VOLTage:AC:RANGe:STEP[:INCRement]
MINimum|MAXimum
minimum|maximum reference level
*RST state:
70 dB
µ
V
Example:
VOLTage:AC:RANGe 90 dBµV
. . . :RANGe?
[MINimum|MAXimum]
Query of reference level
Parameters:
none
query
of
current
reference
level
MINimum|MAXimum
query of lowest|highest reference level
Result:
Reference level in dBµV.
Example:
VOLTage:AC:RANGe? -> 90
. . . . :STEP
. . . . . [:INCRement]
<numeric_value>|MINimum|MAXimum
Setting the stepwidth for the command
VOLTage:AC:RANGe UP|DOWN
Parameters:
<numeric_value>
stepwidth of reference level in dBuV
MINimum|MAXimum
setting the minimum|maximum stepwidth
*RST state:
1 dBuV
Example:
VOLTage:AC:RANGe:STEP 10 dBuV