![background image](http://html2.mh-extra.com/html/rands/eb200/eb200_manual_3079696154.webp)
EB200 Manual
SENSe subsystem MSCan
4052.2000.02 4.80
E-7
. :MSCan
The
MSCan
system controls the frequency function of the device, provided the memory scan has been
activated by
SENSe:FREQuency:MODE MSCan
. Each scan is started by
INITiate[:IMMediate]
.
The memory locations are placed in the
MEMory
subsystem and are set for query during the scan.
. . :CHANnel
<name>
Setting of current memory location.
The memory locations are addressed by the text (see "Parameters" on page 9)
MEM0
to
MEM999
(memory location 0 to memory location 999) and the next free memory location by
ACTUAL
.
During the memory scan, this command is not permitted.
Parameters:
<name>
MEM0
to
MEM999 | NEXT
*RST state:
MEM0
Example:
MSCan:CHANnel MEM357
. . :CHAnel?
Output of current memory location.
Parameters:
none
Result:
Number of current memory location.
Example:
MSCan:CHANnel?
→
357
. . :CONTrol
. . . :[ON]<control_function> {,<control_function>}
Command for switch-on of the
’STOP:SIGNal’
functions
.
With ’
STOP:SIGNal’
the disappearance of the signal during the dwell time causes the dwell time to
be aborted. The hold time after the disappearance of the signal is set with
SENSe:MSCan:HOLD:TIME
.
Parameters:
<control_function>
is the following string:
’STOP:SIGNal’
switches the signal-controlled dwell time on
*RST state
After *RST there is no control function running
Example:
MSCan:CONTrol ’STOP:SIG’