EB200 Manual
SENSe subsystem GCONtrol
4052.2000.02 4.78
E-7
. :GCONtrol
. . [:FIXed|MGC]
<numeric_value>|UP|DOWN|MINimum|MAXimum
Setting of MGC value
Parameters:
<numeric_value>
gain control factor in dBµV
-30 dBµV = no gain control -> maximum sensitivity
110 dBµV = maximum gain control -> minimum sensitivity
UP|DOWN
increase|decrease of the MGC value by the value set in the
command
SENSe:GCONtrol[:FIXed|MGC]:STEP[:INCRement]
.
MINimum|MAXimum
setting the smallest|largest MGC value
*RST state:
50 dBuV
Example:
GCONtrol 50
. . [:FIXed|MGC]?
[MINimum|MAXimum]
Query of the MGC value
Parameters:
none
query of current MGC value
MINimum|MAXimum
query of smallest|largest MGC value
Result:
Gain control
Example:
GCONtrol? -> 50
. . . :STEP
. . . . [:INCRement]
<numeric_value>|MINimum|MAXimum
Setting the stepwidth for the command
SENSe:GCONtrol[:FIXed|MGC] UP|DOWN
Parameters:
<numeric_value>
MGC stepwidth
MINimum|MAXimum
smallest|largest
stepwidth
*RST state:
1 dB
Example:
GCONtrol:STEP 10