Sensor Configuration
R&S
®
NRX
124
User Manual 1178.5566.02 ─ 07
"Linear"
Amplitude averaging
Remote command:
CALCulate<Measurement>[:CHANnel<Channel>]:AVERage:TYPE
Video Bandwidth
Requires a wideband power sensor.
Sets the video filter bandwidth. Reducing the video bandwidth also increases the trig-
ger sensitivity.
Note:
The video bandwidth must never be smaller than the RF bandwidth of the signal.
Remote command:
CALCulate<Measurement>[:CHANnel<Channel>][:POWer]:VBWidth:ENUM
8.4
Range Settings
Access:
"Measurement Settings" > "Primary Sensor Config" > "Range"
"Measurement Settings" > "Secondary Sensor Config" > "Range"
Available for continuous average, burst average, trace, pulse analysis, time gate, time-
slot, statistics measurements.
Some power sensors have only one measurement range, others have two or three
measurement ranges. For details, see the data sheet of the power sensor.
The measurement ranges are also called measurement paths. All available paths are
continuously and simultaneously measured. Adjacent paths overlap by about 6
dB, and
the final measurement result is achieved by appropriately weighting the measurement
results of all paths.