Remote Control Commands
R&S
®
NRX
234
User Manual 1178.5566.02 ─ 07
Suffix:
<Measurement>
.
1 to 8
Measurement
<DirectionalChannel> 1 to 2
1 = forward, 2 = reflection (reverse)
Parameters:
<state>
OFF | ONCE
*RST:
OFF
Manual operation:
See
CALCulate<Measurement>:RELative<DirectionalChannel>:STATe
<state>
Allows you to relate measured power or a power ratio to a reference value. Whether
the power is measured by one power sensor or whether it is a combined value mea-
sured by two power sensors, is set by:
CALCulate<Measurement>:MATH[:EXPRession]
CALCulate<Measurement>[:CHANnel<Channel>]:FEED<Channel>
Alias:
UNIT<Measurement>:POWer:RELative:STATe
CALCulate<Measurement>:RELative<DirectionalChannel>[:MAGNitude]:
AUTO
Suffix:
<Measurement>
.
1 to 8
Measurement
<DirectionalChannel> 1 to 2
1 = primary sensor, 2 = secondary sensor or 1 = forward, 2 =
reflection (reverse)
Parameters:
<state>
OFF | ON | SET
OFF
Absolute power or power ratio
ON
Relative power or power ratio, using the specified reference
value.
SET
Uses the current measurement value as reference value and
enables the relative measurement.
*RST:
OFF
Manual operation:
See
"Forward Relative State, Reflection Relative State"
on page 55
See
Measurement Settings and Results
www.allice.de