Remote Control Commands
R&S
®
NRX
389
User Manual 1178.5566.02 ─ 07
SYSTem:SENSor<Sensor>:REBoot
Reboots the selected power sensor.
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Usage:
Event
SYSTem:SHUTdown
Shuts down the R&S NRX.
Usage:
Event
13.13.3
Firmware Update
If you want to integrate a firmware update function in an application, use
.
Example
You want to update your R&S NRX with the
NRX_18.01.22.02.rsu
file. This file has
a size of 10242884
bytes.
To send the file to the R&S NRX for updating the firmware, your application has to
assemble a memory block containing:
SYST:FWUP <block_data>
The
<block_data>
are definite length arbitrary block data as described in
The size of the file is 10242884. This number has 8 digits. Thus, the
<block_data>
consist of the following:
●
#
●
8
How many digits follow to specify the file size.
●
10242884
Number that specifies the file size.
●
<file_contents>
Contents of the
*.rsu
file, byte-by-byte
●
0x0a
Delimiter
In this example, you write exactly 10242905
bytes to the R&S NRX, for example by
using a 'viWrite()' function.
The 10242905
bytes result from the values of the list above:
System Information and Configuration