Sensor Configuration
R&S
®
NRX
114
User Manual 1178.5566.02 ─ 07
...........................................................................................115
Exclude from Start, Exclude from End
.........................................................................116
Duty Cycle State
Available for continuous average measurements.
Enables or disables the duty cycle correction.
Remote command:
CALCulate<Measurement>[:CHANnel<Channel>]:CORRection:DCYCle:
STATe
Duty Cycle
Available for continuous average measurements.
Sets the duty cycle for measuring pulse-modulated signals. The duty cycle defines the
percentage of one period during which the signal is active. If the duty cycle is enabled,
the R&S NRX takes this percentage into account when calculating the signal pulse
power from the average power.
For thermal power sensors or multipath power sensors, the duty cycle correction is the
only way to determine the power of pulsed signals.
Remote command:
CALCulate<Measurement>[:CHANnel<Channel>]:CORRection:DCYCle[:
VALue]