PROGRAMMING OF VRAE
4 - 27
4.6.6 Enable / Disable Datalog
The user can choose to enable or disable the datalogging
function on each individual sensor. This allows users to
selectively log certain sensor readings of interest.
1. “Enable / Disable Datalog?” is the sixth sub-menu item in
Table 4.5.
2. Press
[Y/+]
key. The display shows all the installed
sensors in the monitor. A previously selected sensor to
be datalogged is marked with an “* ” next to the sensor
name. The cursor is blinking at the first sensor location.
Press
[Y/+]
key to enable the datalogging for the sensor
and
[N/-]
key to disable the datalogging.
CO* SO
2
H
2
S*
LEL pick
OXY
3. Press
[MODE]
key momentarily to move from one
sensor location to the next one. Repeat Step 2 until all
the sensors that needed to be datalogged are selected.
Press and hold
[MODE]
key for 1 second to save the
new sensor selection.
4. The display shows “Save ?” To confirm the new
selection, press the
[Y/+]
key to accept the change. Press
the
[N/-]
key or the
[MODE]
key to discard the change
and return to the first datalog sub-menu.