COMPUTER INTERFACE
5 - 14
user can also enter a time out period so that the backlight
will be turned off automatically after the time out period. If
Auto
is selected, the ambient light sensor will automatically
turn on and off the backlight.
Clock from PC:
This option allows the PC clock to be downloaded to the
VRAE monitor so that the user does not need to manually
set the clock in the VRAE monitor.
Pump Speed:
This option allows the user to choose the high or low pump
speed for the VRAE monitor.
Measure:
The user can choose to record the average or peak (Max)
reading of all the sensors during each datalog period.
Datalog Interval:
The datalog interval is the rate at which datalog samples are
recorded while the datalog is running. The unit of datalog
interval is second. The datalog interval determines how
often a new sample is calculated and stored to the datalog.
Datalog Selection:
The user can choose one or more sensors for datalogging
purpose. This does not activate or inactivate the particular
sensor except for the purpose of datalogging recording.
Datalog Mode:
This may be set to either of four choices:
-- With
Automatic start/stop
, the datalogging begins when
the monitor is turned on and ends when it is turned off.