COMPUTER INTERFACE
5 - 21
5.5.2 Send Configuration to VRAE
After the configuration data is loaded into the ProRAE-Suite
software, the user can send the configuration settings to the
VRAE monitor. To send the configuration settings to the
instrument, click
Communication->Send Config
menu
item or click the send toolbar icon (right arrow with “Send”
letters), a message box will appear to remind the user to
connect the instrument to the PC through the serial port.
After making sure that the instrument is connected to the
PC, click the
OK
button to start the data transfer.
After the configuration information is sent successfully, a
message box will appear to indicate that the sending
configuration process is completed. Click the
OK
button to
close the message box.
If the communication is not successful, an error message
indicating that there is no response from the instrument will
appear. Check the cable to make sure that the connectors on
both ends of the cable are securely seated in the sockets.
Try to send the information again. If the error message still
appears after repeated attempts, call an authorized service
center for help.