PROGRAMMING OF VRAE
4 - 24
4.6.3 Change Datalog Period
The datalog period can be programmed from 1 to 3,600
seconds (1 hour).
1. “Change Datalog Period?” is the third sub-menu item in
Table 4.5.
2. Press the
[Y/+]
key, the display shows “New Period =
0060” with the left-most digit flashing, where “0060” is
the previously stored data log period.
3. To modify this period, starting from the left-most digit,
use the
[Y/+]
or
[N/-]
key to change the digit value and
press the
[MODE]
key momentarily to advance to the
next digit. The flashing digit will move to the next digit
to the right. Repeat this process until all 4 digits of the
new period are entered. Press and hold the
[MODE]
key
for 1 second to exit data entry mode.
4. If there is any change to the existing value, the display
will show “Save?” Press the
[Y/+]
key to accept the new
value. Press the
[N/-]
key to discard the changes and
move to the next menu option.
5. To preserve the previously stored period, press
and hold
the
[MODE]
key for 1 second. The monitor will exit data
entry mode and move to the next sub-menu item in Table
4.5.