Inputs cable label
Outputs cable label
9.7.2. Intrinsic Safety Parameters for IN/OUT Circuits
OUT cables pairs (outputs): 1-2, 3-4, 5-6, 7-8
Ui 30V
Ii any
value
Pi 0.49W
Li negligibly
small
Ci negligibly
small
IN cables pairs (inputs): 1-2, 3-4, 5-6, 7-8
Ui 30V
Ii any
value
Pi any
value
Li negligibly
small
Ci negligibly
small
9-10 OUT cable pair (outputs) and 9-10 IN cable pair (inputs). CIRCUITS
CONNECTED IN PARALLEL
Uo 13.5V
Io 42mA
Po 0.52W
Total Co
0.49µF
Total Lo
0.5mH