48
15.4. Gravitational Coefficient
The function of gravity correction compensates changes of gravity force being
a result of different latitude. It allows correct scale adjustment away from the
point of subsequent use. The value of gravity correction must be entered with
reference to tables prepared by „RADWAG Wagi Elektroniczne” or calculated
using the below formula:
kal.
uzyt.
g
g
Gcor
=
Correction value ranges between 0.90000 ÷ 1.99999.
If the weighing instrument is calibrated in the place of use,
then the value of <GCOR coefficient> parameter must be
1.00000. If the scale is calibrated far from the place of use
(longitudinal change) the value must be corrected.
16. CONNECTION CABLES LIST
PT0327
: scale – IM01 cable.
PT0328
: scale – RS485 cable.
PT0329
: scale – RS232 cable.
17. STANDARDS LIST
The device is manufactured in accordance with the following standards:
1. EN 61326-1:2013
Electrical equipment for measurement, control and
laboratory use - EMC requirements - Part 1: General requirements.
2. EN 61010-1:2010
Safety requirements for electrical equipment for
measurement, control and laboratory use - Part 1: General requirements
.
3. EN 60079-0:2012 + A11:2013
Explosive atmospheres - Part 0: Equipment -
General requirements.
4. EN 60079-11:2012
Explosive atmospheres - Part 11: Equipment protection
by intrinsic safety "i
”.
5. EN 60529:1991 + A2:2013
Degrees of protection provided by enclosures
(IP Code).