44
15.1.2. Factory Parameters List
Name Default
Range
Description
Metrology -
-
Metrology-connected
settings
Mass divisions
-
-
Converter's divisions.
Adjustment unit
kg
g, kg, lb
Adjustment (calibration) unit.
GCOR coefficient
1
0.9
÷
1.1
Gravity correction factor (refer to
section 15.4).
Reading unit - Range 1
0.001
0.0000001
÷
50
Reading units for Range 1.
Verification unit -
Range 1
None
None, 0.001, 0.002,
0.005, 0.01, 0.02,
0.05, 0.1, 0.2, 0.5, 1,
2, 5, 10, 20, 50
Verification units for Range 1,
'none' value - unverified version.
Reading unit - Range 2
0.001
0,0000001
÷
50
Reading units for Range 2.
Verification unit -
Range 2
None
None, 0.001, 0.002,
0.005, 0.01, 0.02,
0.05, 0.1, 0.2, 0.5, 1,
2, 5, 10, 20, 50
Verification units for Range 2,
'none' value - unverified version.
Digits marker
None
None, 1, 2, 3
Setting quantity of digits (counting
from the last digit of the displayed
result) over which a marker is to be
displayed and printed out in case of
unverified balances and scales.
Max range
3.009
-
Maximum range + overload.
Range 2 threshold
0.000 -
II range switching point. For single
range weighing devices set 0.
External adjustment
weight
3.000
-
External adjustment weight mass.
Autozero range
Predefined
Predefined, 0.1, 0.2,
0.25, 0.5, 0.6, 0.7,
0.8, 0.9, 1, 2, 2.5, 3,
4, 5, 6, 7, 8, 9, 10.
Autozero range [d]: Predefined -
value taken from program-
implemented tables; 0.1
÷
10 -
value entered directly by a user.
Autozero range
Predefined
Predefined, 0.1, 0.2,
0.25, 0.5, 0.6, 0.7,
0.8, 0.9, 1, 2, 2.5, 3,
4, 5, 6, 7, 8, 9, 10.
Autozero range [d]: Predefined -
value taken from program-
implemented tables; 0.1
÷
10 -
value entered directly by a user.
Autozero time
Predefined
Predefined, 0, 0.2,
0.4, 0.6, 0.8, 1, 2, 3,
4, 5, 6, 7, 8, 9, 10, 15,
20.
Autozero time [s]: Predefined -
value taken from program-
implemented tables; 0
÷
20 - value
entered directly by a user.
Stability range
Predefined
Predefined, 0.1, 0.2,
0.25, 0.5, 0.6, 0.7,
0.8, 0.9, 1, 2, 2.5, 3,
4, 5, 6, 7, 8, 9, 10.
Stability range [d]: Predefined -
value taken from program-
implemented tables; 0.1
÷
10 -
value entered directly by a user.