38
14.2. Intrinsic Safety Parameters
Intrinsic safety parameters for analog interface, marked with PLATFORM sign,
featuring permanently fixed connecting cable (used to connect load cells):
Uo 5.88V
Io 87mA
Po 0.51W
Co 1µF
Lo 110µH
14.3. 6-Wire Load Cell Connection
Connect 6-wire load cell to the main board following the diagram below:
6-wire load cell connection
466R RADWAG board marking
Load cell signals
Housing SHIELD
REF+ SENSE
+
REF- SENSE
-
IN+
IN- OUTPUT-
+5V INPUT+
AGND INPUT-