37
8. If necessary, cut the wires making them 5cm long (measured from the ferrite
toroid bead).
9. Connect respective wires to load cell's interface (7) on the main board
466R1703.
10. Terminate the cable with a ring terminal of 4.3mm diameter and screw
it to a threaded pin (6) on the indicator's housing.
11. Fit the cable into the cable clamp (4), use manufacturer-delivered cable tie.
12. Fix the cable in the cable gland preventing it from moving, do it by
tightening the gland's cup nut (8), 5Nm torque.
13. Apply the cover and evenly screw the bolts (two turns) fixing the cover,
torque 0,5Nm.
14. Apply security seals.
Load cells selection is conditioned by level of protection
against explosion of the whole device.