10. INSTALLATION
Prior the first use, carefully read this User Manual. Use the device only as
intended. For details concerning 'indicator' - 'load cells' connection, read later
sections of this user manual.
RADWAG does not bear any responsibility for damage or losses resulting
either from improperly carried out installation or misuse.
Prior installation and start, it is necessary to analyse whether
the device complies with the usage requirements regarding
particular hazardous area. The analysis must be carried out by
a qualified personnel.
10.1. Unpacking and Preparing for Operation
•
The indicator must be unpacked in safe area.
•
Prepare
the
workstation.
•
Take the indicator to the workstation.
•
Ground the indicator.
•
Connect the indicator to the mains.
Install the weighing indicator, connect the weighing platform,
and connect the grounding when there is no risk of explosive
atmosphere occurrence.
10.2. Workstation
PUE HX5.EX-* indicator's workstation requirements:
•
Dry, even, horizontal surface.
•
Ambient temperature range: -10°C - +40°C.
•
No heat source in a close vicinity, no risk of intense temperature variation.
•
No mechanical shocks and vibrations.
•
No exposure to mechanical or chemical stress (hazard).
•
The workstation must be located in a considerable distance from
processes where mechanisms causing electrostatic charges greater than
those caused by rubbing the surface by hand, occur.
•
No exposure to sunlight.
•
At the workstation there must be MEB (main equipotential bonding)
installed.