47
•
Upon result stabilization, press
key to confirm. Edit box
<Mass>
is displayed again (second linearity correction point).
•
Linearity correction process for the 2nd point is analogous to the linearity
correction process for the 1st point.
•
Upon declaring requested quantity of linearity correction points and upon
display of the next edit box
<Mass>
, press
key.
•
<Section linearity>
submenu is displayed automatically.
Upon pressing
key in order to confirm zero value for
linearity correction, message: <Value too low> is displayed.
15.3.2. Corrections
Editing corrections values for particular linearity points.
Procedure:
•
Enter
<Factory/Section linearity>
submenu.
•
Select
<Corrections>
parameter. List of determined linearity correction
points is displayed, the list contains previously declared masses,
for particular correction points, and deviations, expressed in scale reading
units.
•
Select requested position, edit box
<Mass>
is displayed.
•
Enter requested correction and/or press
key to confirm. Edit box
<Correction>
is displayed automatically.
•
Enter requested correction and/or press
key to confirm.
<Corrections>
submenu is displayed automatically.
15.3.3. Deleting Linearity
•
Enter
<Factory/Section linearity>
submenu.
•
Select
<Delete>
parameter. Message
<Delete?>
is displayed.
•
Press
key to confirm.