- 30 -
1 Internal adjustment
Pressing this pictogram initiates internal adjustment process. The
process is carried out fully automatically, without operator
intervention. If the weighing pan of a balance is loaded, then the
software displays a command to take the load off the weighing
pan.
2 External adjustment
Pressing the pictogram causes initiating the adjustment process
using an external weight. The value of the external weight is
stored in balance factory memory. The external adjustment
function is not accessible in verified balances. Changing the
adjustment weight value is not possible.
3 User adjustment
Pressing this pictogram causes initiating adjustment process
carried out with an external weight of optional mass. The weight
mass is specified before starting the adjustment process. The user
adjustment function is not accessible in verified balances.
4 Adjustment test
The adjustment test is a comparison of the actual internal weight
value with its value stored in balance memory. The comparison
results in a deviation demonstrating drift of balance indication.
5 Adjustment weight correction
The function enables correcting the value of the internal
adjustment mass. The weight correction function is not accessible
in verified balances.
6 Automatic adjustment
The function enables determining the criteria for initiating the
automatic internal adjustment process. Available options:
no – adjustment disabled
time – adjustment triggered by time interval set in parameter
<Auto. adj. time>
temperature – adjustment triggered by changes of ambient temperature
both – adjustment triggered both by time interval and temperature changes
7 Automatic adjustment time
Determining time interval between the consecutive automatic
adjustment processes – available settings from 1 to 12 hours
8 Adjustment report printout
Determining a criterion whether on completing adjustment or test
the balance should generate and print a report
9 Report setings
Determining the content of a report from adjustment process.
10 Adjustment history
Displaying the history of completed adjustments.
Summary of Contents for MYA.3Y
Page 2: ...2 APRIL 2013...