- 96 -
18.16.1. Select method for comparing mass
The mode enables selecting between two testing methods: ABA and ABBA.
Select name of a required testing procedure.
PROCEDURE:
o
in an optional moement press grey coloured workspace area
o
to open a screen with internal menu of the MASS COMPARISON mode
o
scroll down to
Measuring Series
field
o
press
Measuring Series
field
o
Select a comparing method to be used
o
Select number of method repetitions.
The available number of repetitions ranges from 1 to 20, e.g. 10 x ABA is
in practice 30 measuring processes.
o
Accept seelcted value by pressing OK button.
o
o
Return to weighing:
−
Press screen field with
settings pictogram, or
−
Press “RETURN” arrow
The entered changes are automatically saved. The mass comparator is ready
to use.