- 62 -
18.8.3. ACAI function – Automatic Accuracy Correction
The ACAI function is to automatically correct mass of a single part during
parts counting process. it is aimed to increase the accuracy of counting
process – mass of a single part is calculated from a higher quantity of
samples, ensuring higher accuracy.
First, enable the ACAI function.
PROCEDURE:
o
In an optional moment press
grey coloured workspace area
o
Which opens a screen on
balance internal menu
o
Scroll the menu to option
ACAI
o
Press ACAI button
o
In an opened message box select option YES
Principles of using the ACAI
function:
1. Number of parts (on adding) on
balance’s weighing pan has to
be greater than has been
previously
2. number of parts (on adding) on
balance’s weighing pan must be
less than twice the amount of
which was visible on the display
before adding parts
3. current quantity of parts must be within the
0
value,
4. measurement result has to be stabilized.
Enabled
ACAI
function is indicated
by a black pictogram displayed
next to the number counted of
parts. If information item “Standard
mass” is enabled in the internal
mode menu (access by pressing
the workspace area), then in
meeting the requirements 1-4 the
displayed mass of a single part is
subject to automatic accuracy
correction.
This value can be saved in the “Database of Standards” using a
corresponding quick access key.