G. In case of damage, immediately unplug the device from the
mains.
H. Platform to be decommissioned, should be decommissioned in
accordance with valid legal regulations;
1.6. Temperature Stabilization Time
Before start of the measuring processes, it is necessary to wait until
the platform reaches thermal stabilisation.
Platforms stored in much lower temperatures than the workroom
temperature, before being plugged to the mains (e.g. during winter
period), must be subjected to thermal stabilisation. The thermal
stabilisation period takes about 8 hours. During the thermal
stabilization, the indications on the screen can change. Potential
workroom temperature change shall occur gradually and slowly in the
course of the weighing instrument operation.