47
NB_A_”Serial No.” CR LF
- command understood, response: serial number.
NB_I CR LF
- command understood but not accessible at this
moment.
”Serial No.”
– serial number of the device, Inserted in between
inverted commas.
Example:
NB CR LF
– command sent from a computer.
NB_A_”123456” CR LF
– scale serial number - 123456.
Give accessible units
Format:
UI CR LF
Response options:
UI_”x
1
,x
2
, … x
n
”_OK<CR><LF>
- command carried out, response: accessible
units.
UI_I <CR><LF>
- command understood but not accessible at this
moment.
x
- unit symbol, separated by means of commas.
Example:
UI CR LF
- give available units.
UI_”kg,N,lb,u1,u2”_OK CR LF
- response: available units.
Set current unit
Format:
US_x CR LF
Response options:
US_ x_OK CR LF
- command carried out, response: set unit.
US_E CR LF
- error in-course of command execution, no parameter or
incorrect format.
US_I CR LF
- command understood but not accessible at this moment.
x
- parameter, units symbols: g, kg, N, lb, oz, ct, u1, u2, next.
Caution:
If x=next the command swaps to another available unit on the list (it
simulates
key pressing).
Summary of Contents for HRP 1100
Page 2: ...2 JULY 2018...
Page 4: ...4 1 GENERAL INFORMATION 1 1 Dimensions HRP 16 32 HRP 62 120...
Page 5: ...5 HRP 150 300 HRP 300 1 600 1100...
Page 6: ...6 HRP 2000 1 2 Connectors Arrangement HRP 16 32...
Page 7: ...7 HRP 62 120 HRP 150 300 300 1 600 1100 HRP 2000...
Page 18: ...18 3 Install weighing pan receivers fit them into lever seats...
Page 26: ...26 4 Disassemble transport lock by undoing screws and removing fixing washers 6...
Page 28: ...28 10 Connect the scale to the power supply...
Page 33: ...33 11 Connect the scale to the power supply...
Page 53: ...53 RS232 HRP HY10 PUE 5 cable power supply Ethernet HRP HY10 PUE 5 cables...
Page 54: ...54 Ethernet HRP Ethernet Switch Pue 7 1 cable HRP IN OUT cable...
Page 57: ...57...
Page 58: ...58...