49
Example:
FS CR LF
- give Max capacity.
FS_A_”3.000” CR LF
– Max capacity: „3.000”.
Give program version
Format:
RV CR LF
Response options:
RV_A_”x” CR LF
- command understood, response: program version.
RV_I CR LF
- command understood but not accessible at this moment.
x
– program version (in between inverted commas).
Example:
RV CR LF
- give program version.
RV_A_”1.0.0” CR LF
– response program version „1.0.0”.
Set autozero
Format:
A_n CR LF
Response options:
A_OK CR LF
- command carried out.
A_E CR LF
- error in-course of command execution, no parameter or
incorrect format.
A_I CR LF
- command understood but not accessible at this moment.
n
- parameter, decimal value determining autozero settings:
0 – autozero off.
1 – autozero on.
Example:
A_1 CR LF
- turn autozero function on.
A_OK CR LF
– autozero function is on.
Send all implemented commands
Format:
PC CR LF
Response:
Summary of Contents for HRP 1100
Page 2: ...2 JULY 2018...
Page 4: ...4 1 GENERAL INFORMATION 1 1 Dimensions HRP 16 32 HRP 62 120...
Page 5: ...5 HRP 150 300 HRP 300 1 600 1100...
Page 6: ...6 HRP 2000 1 2 Connectors Arrangement HRP 16 32...
Page 7: ...7 HRP 62 120 HRP 150 300 300 1 600 1100 HRP 2000...
Page 18: ...18 3 Install weighing pan receivers fit them into lever seats...
Page 26: ...26 4 Disassemble transport lock by undoing screws and removing fixing washers 6...
Page 28: ...28 10 Connect the scale to the power supply...
Page 33: ...33 11 Connect the scale to the power supply...
Page 53: ...53 RS232 HRP HY10 PUE 5 cable power supply Ethernet HRP HY10 PUE 5 cables...
Page 54: ...54 Ethernet HRP Ethernet Switch Pue 7 1 cable HRP IN OUT cable...
Page 57: ...57...
Page 58: ...58...