11.13.2. Means of Operation
•
Enter <PIPETTES CALIBRATION.> mode.
•
Start the process.
•
Follow the prompts to set the correct options.
•
Enter the ambient temperature, next press ENTER for confirmation. The computer software
proceeds to the next step, i.e. ambient humidity settings.
•
Set the ambient humidity, next press ENTER for confirmation. The computer software proceeds
to the next step, i.e. air pressure settings.
•
Set the air pressure, next press ENTER for confirmation. The computer software proceeds to
the next step, i.e. entering the first control volume (V1) for a tested pipette.
•
•
Enter (V1) control volume. For fixed volume pipettes (P2.13.5 VOLUME DETERMINATION
parameter set to value <1>) this is the one and only value to be entered. The computer
software proceeds to the next step, i.e. entering the second control volume (V2) for a tested
pipette.
•
Enter (V2) control volume. The computer software proceeds to the next step, i.e. entering the
third control volume (V3) for a tested pipette.
•
Enter (V3) control volume.