- 63 -
•
When all the measurements are saved, it is possible to check the results of statistics by
pressing
button:
•
Select an option <RESULT> by pressing
button.
•
The information on number of saved measurements is displayed in a bottom line. On pressing
or
, the type of displayed information changes.
•
Selecting an option <PRINT> and pressing
button initiates printing out the statistics data
in a form of a report.
An exemplary report:
-------------- Statistics --------------
N
9
Sum
455.600 g
Avg
50.6222 g
Min
49.939 g
Max
51.380 g
Dif
1.441 g
Sdv
0.39605 g
Rdv
0.78 %
----------------------------------------
11.10.2. Deleting Statistics
To delete statistics data carried out for a series of measurements, follow this procedure:
Summary of Contents for AS R2
Page 35: ...35...
Page 70: ...70 An exemplary report adjustable volume pipette 3 volumes tested...
Page 95: ...95...