- 36 -
The measurements saved in the balance’s memory cannot be deleted.
It is possible for a user to view and print out the data saved in the memory.
Procedure:
Each measurement is saved with its individual number. The format is: b4.4.n, where <n> is the
consecutive number of the saved measurement. In the bottom line the date and time are displayed
for every single measurement.
To swap between measurements recorded in database, use the arrow buttons,
or
.
Pressing one of these buttons lets the user move to the next measurement either up or down the list.
To view the remaining data relating to a respective measurement, first select the measurement in
question, next press
button:
The software automatically shows the data relating to the measurement, data is presented in the
bottom line of the balance display.
To swap between data relating to the measurement, press
or
. The data relating to the
measurement can be printed out by selecting the option <PRINT> and pressing
button.
An exemplary printout:
Date
21.06.2013
Time
13:05:02
User
Product
Tare
0.000 g
Gross
0.000 g
0.000 g
-----------------Cal. report.---------------
Cal. Type
Internal
User
Project
1234567890123459
Date
16.07.2013
Time
13:27:09
Balance ID
10353870
Cal. diff.
-0.004 g
Summary of Contents for AS R2
Page 35: ...35...
Page 70: ...70 An exemplary report adjustable volume pipette 3 volumes tested...
Page 95: ...95...
Page 96: ...96...