- 92 -
Set ambient conditions state
Format: EV_n CR LF
Response options:
EV_OK CR LF
– command carried out
EV_E CR LF
– error in-course of command execution, no parameter or incorrect format
EV_I CR LF
– command understood but not accessible at this moment
n
– parameter, decimal value determining ambient conditions state
n
→
0 – unstable ambient conditions
1 – stable ambient conditions
CAUTION!
Command changes settings for a current working mode.
An example:
Command:
EV_1 CR LF
– set value ‘stable’ for ambient conditions option
Response:
EV_OK CR LF
– ambient conditions option set to value ‘stable’
<AMBIENT CONDITIONS> parameter is set to value <STABLE> until command EV 0 swaps it to
value <UNSTABLE>.
Set filter
Format: FIS_n CR LF
Response options:
FIS_OK CR LF
– command carried out
FIS_E CR LF
– error in-course of command execution, no parameter or incorrect format
FIS_I CR LF
– command understood but not accessible at this moment
n
– parameter, decimal value determining filter number
n →
1 – very fast
2 – fast
3 – average
4 – slow
5 – very slow
CAUTION!
The numbering is assigned to a particular filter name and it is identical for all balance types.
The command changes settings for a current working mode if, for a particular balance type, filter settings are
assigned to the working mode.
An example:
Command:
FIS_3 CR LF
– set average filter
Response:
FIS_OK CR LF
– average filter set
Set Value release
Format: ARS_n CR LF
Response options:
ARS_OK CR LF
– command carried out
ARS_E CR LF
– error in-course of command execution, no parameter or incorrect format
ARS_I CR LF
– command understood but not accessible at this moment
n
– parameter, decimal value determining value release options
n →
1 – fast
2 – fast+reliable
3 – reliable
CAUTION!
The numbering is assigned to a particular value release option and it is identical for all balance types.
The command changes settings for a current working mode if, for a particular balance type, value release
settings are assigned to the working mode
An example:
Command:
ARS_2 CR LF
– set value release parameter to fast+reliable option
Response:
ARS_OK CR LF
– fast+reliable option set
Summary of Contents for AS R2
Page 35: ...35...
Page 70: ...70 An exemplary report adjustable volume pipette 3 volumes tested...