Luna  

LAB, LPB, LTB, LBB Series 

 

Operating Manual 

For internal (‘i’) and external (‘e’) calibration models 

 

 
Software rev.: 

V1.0105 & above (Analytical Models) 

 

 

 

V2.0104 & above (Precision Models) 

 
 

          
        

 

Summary of Contents for Luna LAB 124 e

Page 1: ...quipment Company 2021 Adam Equipment Luna LAB LPB LTB LBB Series Operating Manual For internal i and external e calibration models Software rev V1 0105 above Analytical Models V2 0104 above Precision...

Page 2: ...Adam Equipment Company 2021...

Page 3: ...6 10 2 PASSCODES 16 10 3 WEIGHING 17 10 4 FUNCTIONS 18 10 4 1 Parts Counting 19 10 4 2 Percentage Weighing 19 10 4 3 Check weighing 20 10 4 4 Animal Dynamic Weighing 21 10 4 5 Net Total 23 10 4 6 Dens...

Page 4: ...Adam Equipment Company 2021 2 17 0 SERVICE INFORMATION 38 18 0 BALANCE MENU STRUCTURE 39 19 0 LANGUAGE TABLE 42 20 0 WARRANTY INFORMATION 43...

Page 5: ...ales service information and other important information These balances are highly accurate precision instruments and contain sensitive mechanisms and components They should be transported and handled...

Page 6: ...ications include weighing percentage weighing parts counting dynamic animal weighing and solid and liquid density determination Bi directional RS 232 interface and USB interface as standard Can be con...

Page 7: ...T Taels S Mommes Tolas Ticals pennyweight custom Stabilization Time Typically 3 seconds Operating Temp 15 C to 35 C recommended 40 60 RH non condensing Power Supply External mains power adapter suppli...

Page 8: ...xternal mains power adapter supplied as standard Input Voltage 100 240 VAC 50 60 Hz Removed text related to battery Input Voltage 18 VDC 830 mA Weighing mechanism Precision Load Cell Calibration Suffi...

Page 9: ...H non condensing Power Supply External mains power adapter supplied as standard Input Voltage 100 240 VAC 50 60 Hz Input Voltage 18 VDC 830 mA Weighing mechanism Force Restoration Calibration Suffix i...

Page 10: ...mp 15 C to 35 C recommended 40 60 RH non condensing Power Supply External mains power adapter supplied as standard Input Voltage 100 240 VAC 50 60 Hz Input Voltage 18 VDC 830 mA Weighing mechanism Pre...

Page 11: ...tion Time Typically 3 seconds Operating Temp 15 C to 35 C recommended 40 60 RH non condensing Power Supply External mains power adapter supplied as standard Input Voltage 100 240 VAC 50 60 Hz Removed...

Page 12: ...in direct sunlight or near air conditioning vents Avoid unsuitable tables The table or floor must be rigid and not vibrate Avoid unstable power sources Do not use near large users of electricity such...

Page 13: ...s and 12 hours for LAB LPB models 6 1 3 Weighing Insert the power supply cable DC connector to the connector on the rear of the balance Plug the power supply module into the mains and press the power...

Page 14: ...to select calibration weight value and then press the Setup display shows xxxx g load the weight Display shows Display shows remove the weight and clear the top pan Display shows then returns to zero...

Page 15: ...e display show the current operation or function being used 7 1 SYMBOLS AND TEXT The LCD has unique symbols to indicate the following 0 Zero Busy Stable g mg kg ct dwt GN ozt oz N Custom g cc Pcs Symb...

Page 16: ...e balance to print data Mode Mode Enters the Mode Selection Menu Unit Unit Selects weighing units by cycling through a set of enabled units Setup Setup Enters the setup parameters Supervisor Menus Ent...

Page 17: ...press Up and Down to increase or decrease the flashing digit as required Once the value displayed on screen is as required press the Setup key to accept or enter the displayed value Press the 0 T key...

Page 18: ...10 2 PASSCODES This equipment has passcode security functions which can restrict certain operations to particular users Supervisor and Operator modes are available If no passcode is set then the defau...

Page 19: ...change to show the weight in the selected weighing unit The available weighing units can be enabled or disabled during setup of the balance see section 13 1 Only weighing units that have been enabled...

Page 20: ...20000 1 kg 0 420000 0 000001 ct 2100 0 005 dwt 270 066 0 001 GN 6481 59 0 02 ozt 13 50330 0 00005 oz 14 81505 0 00005 Lb 0 92594 0 00001 N 4 1188 0 0001 10 4 FUNCTIONS When weighing the user can acces...

Page 21: ...number of parts based on the average piece weight The display will show the result in number of pieces This will be an integer value in the format XX PCS To count another item press Mode and continue...

Page 22: ...splayed to the maximum number of decimal places based on the resolution of the balance To increase or decrease by one decimal place press the Up or Down key respectively 10 4 3 Check weighing Press Mo...

Page 23: ...he balance will display for 1 second then show the current average weight with the Animal symbol flashing To resume press Print again or if you do not wish to continue then pressing Mode will display...

Page 24: ...d value press Setup or to escape without changing the value press Mode 10 4 4 3 2 Test time Press Setup when is shown and the display will next show the number of seconds over which the balance will a...

Page 25: ...rse the sample in the chosen liquid The density of the liquid must be known or determined from a look up table Steps Press Mode and then Up and Down keys until Density Solid or Density Liquid symbol i...

Page 26: ...the density determination kit the volume of the plumb is marked on its support e g 10 123 cc Steps Press Mode and then Up and Down keys until Density Liquid symbol is displayed and then press Setup t...

Page 27: ...n 2 input to balance RXD Pin 3 output from balance TXD Pin 5 Signal ground GND Handshaking is not applied Baud rate Selectable 4800 9600 19200 38400 Parity Selectable NONE 8N1 EVEN 7E1 or ODD 7O1 All...

Page 28: ...ghing 12 345 Density 12 345 g cc 11 2 2 Custom output format If output on demand is selected the user may optionally configure the serial output as a choice of 3 styles of form either in a default for...

Page 29: ...ghing mode KS CR Enters the setup section This is the same as pressing the Setup key when the balance is in the normal weighing mode Once entered the setup section the balance can be controlled remote...

Page 30: ...ing within the limited parameters The errors likely to occur are A D counts below lowest allowed value A D counts above highest allowed value A D not operating Maximum capacity exceeded Other errors m...

Page 31: ...press Setup The right hand side of the display will show the symbol for the first unit e g carats ct together with its enable state or The Supervisor can then enable or disable the carats unit by usi...

Page 32: ...ed the format be changed by the user using a selection of available data By default the 2 forms are the same as the standard form unless changed by the user as below When 1 or 2 is selected the user c...

Page 33: ...On Enable sound alerts OFF Disable sound alerts AUTO Always on unless balance is not used for 5 minutes then turns off automatically until key is pressed or weight 20d is detected ON Permanently on te...

Page 34: ...ress Mode to advance to setting of the next menu or 0 T to return to normal weighing 13 7 PASSCODES To enable the security features in this balance it is necessary to set passcodes There are 2 passcod...

Page 35: ...determine the density of the liquid 14 2 ANTI VIBRATION TABLE The anti vibration table is a support for laboratory balances that isolate the balance from vibration through the floor The table has a gr...

Page 36: ...also speak the readings received This is ideal if you want to stay informed of a scale s progress whilst completing other tasks or maybe you might be visually impaired See http www adamdu com for fur...

Page 37: ...parameter to a normal value will restore operation Mechanical Problems The balances consist of complicated and fragile mechanical devices They can be damaged by placing a weight on it which is too hi...

Page 38: ...ity on the samples can cause drifting and instability Check the area around the weighing pan for hair dust obstructions under the pan A complete inspection of the mechanics to look for sources of fric...

Page 39: ...Make sure all mechanical problems have been eliminated first OTHER PROBLEMS Cannot calibrate Zero shifted more than allowed Calibration timeout Check all flexures for damage Reset dealer calibration V...

Page 40: ...tails of the unit purchased This part of information should always be available for any future correspondence We suggest you to fill in this form as soon as the unit is received and keep a print out i...

Page 41: ...Adam Equipment Company 2021 39 18 0 BALANCE MENU STRUCTURE Operator Level Access LUNA OPERATING SOFTWARE MENU STRUCTURE...

Page 42: ...options are not available or extra options may be visible depending on if the model is internal or external calibration type and whether it is an approved model Conditions of approval in some countrie...

Page 43: ...Adam Equipment Company 2021 41 Supervisor Level Access continued...

Page 44: ...R Day M O DE S M O DO S M O Dv S m o d e s Modes T Hv R SDA J v E v E S DO NNE R S J E v DI Thursday v NI T S v NI DADEE I NHE I T v n i t e s Units F R I DAY V I E R NE S F R E I T AG V E NDR E D Fri...

Page 45: ...installation unauthorised modifications or attempted repair or failure to observe the requirements and recommendations as given in this User Manual This product may include a rechargeable battery that...

Page 46: ...Adam Equipment Company 2021...

Page 47: ...ercial environment This equipment generates uses and can radiate radio frequency energy and if not installed and used in accordance with the instruction manual may cause harmful interference to radio...

Page 48: ...24242 Felde Germany Phone 49 0 4340 40300 0 Fax 49 0 4340 40300 20 e mail vertrieb aeadam de Adam Equipment S A Pty Ltd 7 Megawatt Road Spartan EXT 22 Kempton Park Johannesburg Republic of South Afric...