fluorescent lamp with household waste. Contact a local waste disposal
for the nearest fluorescent lamp disposal site.
•
QUALIFIED SERVICE –
All service must be performed by the factory or
by a certified CR Pro field service engineer. There are no user replaceable
parts inside the CR Pro.
DO NOT
remove the covers as internal parts may be
damaged, the warranty will be voided and it may be hazardous to your health.
•
SPACE REQUIREMENTS –
The CR Pro has dimensions of
22in(56.4cm)(w) x 21in( 53.9cm)(d) x 40in(102.6cm) (h). An additional 36 in
(92.3 cm) of space should surround the CR Pro to provide access and
working space for documents and plate cassettes.
•
POWER REQUIREMENTS –
Domestic (U.S.): AC outlet 100V – 120VAC 15 amp, 60 Hz
European: AC outlet 220V – 240VAC 10 amp, 47Hz – 63Hz.
The AC outlet should be dedicated to the CR Pro (no other devices
connected to this circuit) to prevent the possibility of injected line noise by
other devices.
The CR Pro is fused for 5 amp domestic and 3 amp European. Fuses are
located in the AC power input fixture on the back, bottom, right hand side of
the CR Pro.
●
FUSE –
The input power is fused at the AC power input connector at the
bottom, right rear of the CR Pro cabinet. The fuse(s) is located in a box
directly above the input connector and is a integral part of the connector
assembly. to remove the fuse box, perform the following:
1. Turn off power to the device.
2. Remove the AC line cord from the connector.
3. Using a small flat tip device, insert the tip under the tab at the bottom
of the box above the connector and gently pry the box open.
4. Inside the box are two (2) fuses as follows:
5. Domestic 100 – 120VAC 5V Fuse(s) (standard)
European 220 – 240VAC 3V Fuse(s) (standard)
Replace the fuse(s) with the value required for your model CR Pro
scanner and replace the box in its receptacle making sure the box is fully
inserted.
Replace the AC line code and turn on power. If the fuse blows again,
6