RM-111 Operations Manual
Application 1
Application 1
27
averaged value in the results.
11. The test is now complete. To process the results press
ENTER
.
12. To re-run the test select
Re-Run Test
. The pulses will be counted again and the new count will
be displayed on the RM-111. Press
ENTER
to continue.
13. To display the calculated results select
Show Results
. The end count is displayed along with a
down arrow icon in the lower right corner. This icon references the
DOWN ARROW
key on the
RM-111 hand controller. By pressing the
DOWN ARROW
key the RM-111 displays the calculated
results. Using the
UP/DOWN ARROW
keys the user can move between the watthour count and
the calculated results.
14. To print the results connect the RM-RPP Remote Portable Printer to the
SERIAL PORT
on the
RM-111. Turn the RM-RPP to the ON 1 position. Select
Print Results
. This will start the printing
process. The printout is time/date stamped and includes both the watthour count and the calculated
% Registration or % Error.
15. To save the results select
Save Results
. The RM-111 will prompt the user for the enabled
parameters in the
User Prompts
menu and save the test results.