RM-111 Operations Manual
12. In this mode, the RM-111 counts the number of pulses from both the reference standard(s)
and the sensor signal from the meter under test. While the test is in progress, the RM-111 display
shows the meter registration as it accrues and also shows the the standards counting. When using
a pulse sensor, the test automatically stops when the pulse sensor senses the last pulse.
13. The test is now complete. To process the results press ENTER.
14. To re-run the test select Re-Run Test. The pulses will be counted again and the new count will
be displayed on the RM-111. Press ENTER to continue.
15. To display the calculated results select Show Results. The end count is displayed along with
a down arrow icon in the lower right corner. This icon references the DOWN ARROW key on
the RM-111 hand controller. By pressing the DOWN ARROW key the RM-111 displays the
calculated results. Using the UP/DOWN ARROW keys the user can move between the watthour
count and the calculated results.
16. To print the results connect the RM-RPP Remote Portable Printer to the SERIAL PORT on the
RM-111. Turn the RM-RPP to the ON 1 position. Select Print Results. This will start the printing
process. The printout is time/date stamped and includes both the watthour count and the calculated
% Registration or % Error.
17. To save the results select Save Results. The RM-111 will prompt the user for the enabled
parameters in the User Prompts menu and save the test results. The saved results also include the
time and date stamp, and both the calculated results and the accumulated watthours during the
test.
Application 9
Application 9
72