Maintenance and Performance Checks 5-9
Gated - External signal at the TRIG IN connector enables the
Model 3152 output. The last cycle of the output waveform is
always completed.
Burst - Each transition at the front panel TRIG IN connector
stimulates the Model 3152 to generate a burst of pre-selected
number of cycles.
Equipment: Pulse/function generator (8021), oscilloscope.
Trigger
1.
Configure the Model 3152 as follows:
Function:
Sine wave
Frequency:
1.000 MHz
Operating Mode: Triggered
Remote Commands:
:RES
:OUTP ON
:FREQ 1e6
:INIT:CONT
OFF
2.
Set the external pulse/function generator
frequency to 10 KHz and connect its SYNC
connector. Set the oscilloscope and verify on the
oscilloscope that the Model 3152 outputs a
triggered signal. Leave the external pulse
generator connected to the Model 3152 for the
next test.
Gate
1.
Configure the Model 3152 as follows:
Function:
Sine wave
Frequency:
1.000 KHz
Operating Mode:
Gated
Remote Commands:
:RES
:OUTP ON
:FREQ 1e3
:INIT:CONT OFF
:GATE:STAT ON
2.
Set the oscilloscope and verify that the Model 3152
outputs a gated signal. Leave the external pulse
generator connected to the Model 3152 for the
next test.
Summary of Contents for 3152
Page 16: ...Getting Started 1 7 Figure 1 1 Segment 1 Sin x x Waveform Figure 1 2 Segment 2 Sine Waveform...
Page 25: ...Configuring The Instrument 2 6 This page was intentionally left blank...
Page 63: ...Using The Instrument 3 38 This page was intentionally left blank...
Page 80: ...SCPI Command Reference 4 17 Figure 4 1 SCPI Status Registers...
Page 121: ...Specifications A 12 This page was intentionally left blank...