3151 And 3151A User Manual
Using The Instrument 3-38
Figure 3-8, Sequenced Waveforms - Triggered Advance Mode
Triggered Sequence
Advance Commands
Placing the Model 3151/3151A in Triggered Sequence Advance
mode is done in Triggered mode only. First, prepare the sequence of
waveforms using the commands that were explained before.
Second, place the instrument in Triggered mode using the
INIT:CONT OFF command. To place the Model 3151/3151A in
Triggered Sequence Advance mode, use the following command:
SEQuence:ADVance {AUTO | TRIGger}
AUTO specifies the normal continuous advance and TRIGger places
the instrument in Triggered Sequence Advance mode.
The query:
SEQuence:ADVance?
queries Triggered Sequence Advance mode and returns AUTO or
TRIG.