![background image](http://html2.mh-extra.com/html/racal-instruments/3151/3151_user-manual_3124279066.webp)
3151 And 3151A User Manual
Using The Instrument 3-36
The byte ordering of the Segment Number section is: High
byte, low byte. There is an offset between the Segment
Number above and the actual segment number. The above
structure ranges from 0 to 4095. The corresponding
segments are selected as ranging from 1 through 4096 with
the TRAC:SEL command.
The Repeat Number binary structure requires restructuring of
the byte order. The MSB (Most Significant Byte) of the 3
bytes is shifted around to the beginning and the mid-byte and
LSB (Least Significant Byte) are shifted left by one byte.
Figure 3-7, Fast Sequence Download
A simple example using the sequence command is to generate a
1 step sequence with segment 1 repeated 74565 times:
SEQ:DATA#15<0000234501h>
where the 15 means:
1 specifies that one more byte will follow
"#"
non-zero
ASCII digit
ASCII digit
Start of
Data Block
Number of Digits
to Follow in header
Byte Count: 5 x Number
of Sequence Steps
Segment Number for
this step: 2 bytes
Number of Repeats for
this step: 3 bytes
restructured as follows:
Binary Number
0-4095 range
Binary # of Repeats:
MSB shifted around to
LSB position
MSB
LSB
Mid
Byte
MSB
Header section (ASCII)
Binary section (binary)
Summary of Contents for 3151
Page 4: ...This page was left intentionally blank...
Page 24: ...3151 And 3151A User Manual Getting Started 1 12 This page was left intentionally blank...
Page 29: ...3151 And 3151A User Manual Configuring The Instrument 2 5 Figure 2 1 Set The Logical Address...
Page 74: ...3151 And 3151A User Manual Using The Instrument 3 44 This page was left intentionally blank...
Page 98: ...3151 And 3151A User Manual SCPI Command Reference 4 24 This page was left intentionally blank...
Page 127: ...3151 And 3151A User Manual Optional Harness Assemblies 6 3...
Page 131: ...3151 And 3151A User Manual Optional Harness Assemblies 6 7...