3151 And 3151A User Manual
Maintenance and Performance Checks 5-7
Amplitude Accuracy
Accuracy Specifications (1 kHz):
"
(1% of r200
µ
V), to 160 mV
"
(1% of r2 mV), to 1.6 V
"
(1% of r20 mV), to 16 V
Equipment: DMM, 50
Ω
feedthrough termination
1. Configure the Model 3151/3151A as follows:
Function:
Sine
Frequency:
1.000 kHz
Amplitude:
15 V
Remote Commands:
:RES
:OUTP ON
:FUNC:SHAP SIN
:FREQ 1000
:VOLT 15
2. Set the DMM to ACV measurements (RMS).
3. Connect the Model 3151/3151A output to DMM input. Terminate
the output with a 50
Ω
feedthrough termination. Set the
amplitude and verify DMM reading as follows:
3151 SETTING
DMM READING
15.0 V
5.373 V - 5.230 V
1.50 V
537.3 mV - 523.0 mV
150 mV
53.73 mV - 52.30 mV