Maintenance and Performance Checks 5-13
Burst
1. Configure the Model 3151/3151A as follows:
Function:
Sine wave
Frequency:
1.000 kHz
Operating Mode:
Burst
Remote Commands:
:RES
:OUTP ON
:FREQ 1e3
:INIT:CONT OFF
:TRIG:BURST:STAT ON
:TRIG:COUN 10
:TRIG:SOUR:ADV:EXT
2. Set the oscilloscope and verify that the Model 3151/3151A
outputs a burst of 10 complete output waveforms.
Remove the external pulse generator connection from
the Model 3151/3151A for the next test.